Integrity, truth-seeking and innovation ──────
Pressure Measurement
Hygienic Pressure Transmitter
Features
■Wide range coverage
■All stainless steel structure, various types of pressure interface, with flush type, hygienic, anticorrosion and other forms
■The output signal is various in form and can be adjusted on site.
■ Reverse polarity protection and transient overcurrent and over voltage protection
GMP501 Series Pressure Transmitter
Model
GMP501-F
Medium
Gas/liquid
Range
-0.1MPa...0MPa~0.01Mpa...100MPa
Over pressure
1.5 times full scale pressure or 110 MPa (minimum value)
Pressure type
Gauge pressure, absolute pressure, sealed gauge type
Precision
±0.25% FS (typical); ±0.5% FS (maximum)
Long-term stability
±0.1%FS/year (typical); ±0.2%FS/year (maximum)
Temperature error
Zero temperature error: ±0.03% FS/°C (≤100kPa); ±0.02%FS/°C (>100kPa)
Full temperature error: ±0.03% FS/°C (≤100kPa); ±0.02%FS/°C (>100kPa)
Operating temperature
‘-30℃~80℃;-10℃~60℃ (explosion-proof type)
Storage temperature
-40℃~120℃
Power supply
15~28V DC
Output signal
4mA ~ 20mA DC (two-wire system) 0mA ~ 10 / 20mA DC (three-wire system) 0 / 1V ~ 5 / 10V DC (three-wire system)
Load Resistance
≤(U-15)/0.02(Ω)(4mA~20mA DC) ≤(U-15)/0.02(Ω)( 0mA~10/20mA DC) >100kΩ(0/1V~5/10V DC)
Protection level
IP65
Electrical connections
Connector or cable 1.5m
Shell material
304
Diaphragm material
316L
O-ring
Fluororubber
Rubber casing
Nitrile rubber
Remarks
Installation methods, body materials, output signals, accuracy requirements, temperature ranges and other parameters can be customized according to different customer needs.